The final rule for implementation of a unique device identifier or UDI has been released and will be posted on the following FDA page on 9-24-13.  Additionally, the FDA has posted the Global Unique Device Identification Database (GUDID) draft guidance available here.

The FDA has provided a summary of compliance dates for the UDI final rule.

1 year after publication – The labels and packages of class III medical devices and devices licensed under the Public Health Service Act (PHS Act) must bear a UDI. § 801.20.
Dates on the labels of these devices must be formatted as required by § 801.18. Data for these devices must be submitted to the GUDID database. § 830.300.

2 years after publication – The labels and packages of implantable, life-supporting, and life-sustaining devices must bear a UDI. § 801.20. Dates on the labels of these devices must be formatted as required by § 801.18.  A device that is a life-supporting or life-sustaining device that is required to be labeled with a UDI must a bear UDI as a permanent marking on the device itself if the device is intended to be used more than once and intended to be reprocessed before each use. § 801.45.
Stand-alone software that is a life-supporting or life-sustaining device must provide its UDI as required by § 801.50(b). Data for implantable, life-supporting, and life-sustaining devices that are required to be labeled with a UDI must be submitted to the GUDID database. § 830.300.

3 years after publication – Class III devices required to be labeled with a UDI must bear a UDI as a permanent marking on the device itself if the device is a device intended to be used more than once and intended to be reprocessed before each use. § 801.45.  The labels and packages of class II medical devices must bear a UDI. § 801.20. Dates on the labels of these devices must be formatted as required by § 801.18.  Class II stand-alone software must provide its UDI as required by § 801.50(b).Data for class II devices that are required to be labeled with a UDI must be submitted to the GUDID database. § 830.300.

5 years after publication – A class II device that is required to be labeled with a UDI must bear a UDI as a permanent marking on the device itself if the device is a device intended to be used more than once and intended to be reprocessed before each use. § 801.45.  The labels and packages of class I medical devices and devices that have not been classified into class I, class II, or class III must bear a UDI. § 801.20. Dates on the labels of all devices, including devices that have been excepted from UDI labeling requirements, must be formatted as required by § 801.18.  Data for class I devices and devices that have not been classified into class I, class II, or class III that are required to be labeled with a UDI must be submitted to the GUDID database. § 830.300.  Class I stand-alone software must provide its UDI as required by § 801.50(b).

7 years after publication – Class I devices, and devices that have not been classified into class I, class II, or class III that are required to be labeled with a UDI, must a bear UDI as a permanent marking on the device itself if the device is a device intended to be used more than once and intended to be reprocessed before each use. § 801.45.

Previous User Requirements
Next Training Records and User Authorizations